正常脑组织磁共振两b值及多b值扩散加权成像技术变异度对比分析
Variation analysis of diffusion weighted imaging in normal brain tissue between two b and multi-b value sequence

李祥;高朋瑞;张首宁;张宏凯;曲金荣;黎海亮;

摘要(Abstract):

目的对比评估两b值及多b值扩散加权成像(diffusion weighted imaging,DWI)技术在正常脑组织中的变异度。材料与方法对29名正常脑组织行3.0 T GE磁共振两b值(0、1000 s/mm~2)及8b值(0、50 s/mm~2、100 s/mm~2、200 s/mm~2、400 s/mm~2、600 s/mm2、800 s/mm~2、1200 s/mm~2)DWI序列扫描,获取两b值参数表观扩散系数(apparent diffusion coefficient,ADC)值及多b值参数包括慢速ADC(D)、快速ADC(D*)、灌注相关体积分数ADC(f),应用变异系数评价脑组织各参数。结果脑组织的ADC值、D值、D*值、f值及其变异系数分别为:脑白质:(81.11±7.62)×10~(-5) mm/s,0.094;(78.47±6.72)×10~(-5) mm/s,0.086;(722.7±189.3)×10~(-5) mm/s,0.262;(8.97±3.63)%,0.404;脑皮质:(87.03±9.97)×10~(-5) mm/s,0.115;(81.83±7.51)×10~(-5) mm/s,0.092;(611.9±169.5)×10~(-5) mm/s,0.292;(9.55±3.41)%,0.357;基底节:(76.70±7.17)×10~(-5) mm/s,0.093;(73.72±6.40)×10~(-5) mm/s,0.087;(1221.9±743.4)×10~(-5) mm/s,0.608;(9.72±3.68)%,0.379。正常脑组织的ADC值及D值变异度接近且较低,而D*值及f值的变异度较高。结论多b值DWI技术获得的D值与两b值DWI技术获得的ADC值在正常脑组织具有同样低的变异度,而D*及f变异度较高,进行临床评价时结果需谨慎解读。

关键词(KeyWords): 磁共振成像;扩散加权成像;变异系数;脑

Abstract:

Keywords:

基金项目(Foundation): 河南省科技计划项目(编号:142102310435)~~

作者(Author): 李祥;高朋瑞;张首宁;张宏凯;曲金荣;黎海亮;